CONTROL COMMANDS

CONTROL COMMANDS

 

Command Message

Meaning

 

 

”C ZERO<CRLF>”

Do a zero check

 

 

”C COMPUTE ZERO<CRLF>”

Calibrate Zero point1

 

 

”C SPAN<CRLF>”

Do a span check

 

 

”C COMPUTE SPAN<CRLF>”

Calibrate Span point1

 

 

”C ASEQ1<CRLF>”

Do a auto-cal sequence2

 

 

”C ASEQ2<CRLF>”

Do a auto-cal sequence2

 

 

”C ASEQ3<CRLF>”

Do a auto-cal sequence2

 

 

”C ABORT<CRLF>”

Aborts auto-cal sequence

 

 

”C EXITZ<CRLF>”

Exit zero cal only

 

 

”C EXITS<CRLF>”

Exit span cal only

 

 

”C EXIT<CRLF>”

Exit zero span or hold

 

1 Executed only if the instrument is in the proper calibration mode and concentration

is within calibration limits. This command adjusts slope and offset values.

2 Initiated only If automatic calibration sequence setup is programmed and enabled.

TABLE 7.4

NOTE

THE COMMANDS IN TABLE 7.4 CAN ONLY

BE ENTERED VIA THE RS-232 PORT WHEN

THE ANALYZER IS IN THE SAMPLE MODE.

When a control command is issued, the CPU will respond by issuing a status report. For example if the host computer issues the command

"C CALZ<CRLF>"

to do a zero check, the CPU will send the status report

"C DDD:HH:MM IIII START ZERO CALIBRATION<CRLF>"

to the RS-232 output.

7.4 Diagnostics

The diagnostics mode can be entered from the RS-232 port as well as from the front panel. The diagnostics commands available are listed below.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 65

Page 71
Image 71
Teledyne 300 instruction manual Diagnostics, Control Commands, Command Message, Meaning

FAQ

How do I perform a zero check?

To perform a zero check, enter the command 'C ZERO&lt;CRLF&gt;' via the RS-232 port while the analyzer is in sample mode.

What is the purpose of the diagnostics mode?

The diagnostics mode allows users to access additional commands for troubleshooting and maintenance.

Can I calibrate the span using the RS-232 port?

Yes, you can calibrate the span by issuing the 'C SPAN&lt;CRLF&gt;' command via the RS-232 port while the analyzer is in sample mode.