DIAGNOSTIC COMMANDS

NOTE

THE DIAGNOSTICS MODE MAY ONLY BE

ENTERED VIA THE RS-232 PORT WHEN

THE ANALYZER IS IN SAMPLE MODE.

DIAGNOSTIC COMMANDS

 

 

 

Command

 

Function

D ENTER SIG

 

Enter diagnostic Signal I/O mode

D EXIT

 

Exit diagnostics mode

D LIST

 

Prints all Signal I/O values. See Table

 

 

5.2 for Signal Definitions

D name[=value]

 

Examines or sets I/O signal. See

 

 

Table 5.2 for a list of signals. Must

 

 

issue D ENTER SIG before using this

 

 

command.

D RESET

 

Reset analyzer(same as power on)

D RESET RAM

 

Resets analyzer and erases RAM.

 

 

Erases all DAS data. Keeps setup

 

 

variables and calibration

D RESET EEPROM

 

Resets analyzer and erases RAM and

 

 

EEPROM. Erases all DAS data.

 

 

Resets all setup variables to factory

 

 

default. Resets calibration values.

 

TABLE 7.5

These commands may be used whether the diagnostics have been entered from the keyboard (SETUP-DIAG) or the RS-232 ("D ENTER <CRLF>"). However, when the diagnostics are entered via the keyboard, no feedback is sent to the RS-232 channel. This prevents the RS-232 output from getting unnecessarily cluttered with diagnostic data.

Whenever the diagnostic mode is entered or exited, a report is issued to the RS-232 output. The table below summarizes the diagnostic reports.

DIAGNOSTIC REPORTS

Report

"C DDD:HH:MM IIII ENTER DIAGNOSTIC MODE"

"C DDD:HH:MM IIII EXIT DIAGNOSTIC MODE"

TABLE 7.6

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 66

Page 72
Image 72
Teledyne 300 instruction manual Diagnostic Commands, DIAGNOSTIC REPORTS Report, Function

FAQ

What is the purpose of the diagnostic commands?

The diagnostic commands allow the user to enter diagnostic Signal I/O mode, print Signal I/O values, reset the analyzer, and more.

How can I enter the diagnostic mode?

The diagnostic mode may only be entered via the RS-232 port when the analyzer is in sample mode. See the manual for specific instructions.

What feedback is sent to the RS-232 channel when entering the diagnostic mode via the keyboard?

No feedback is sent to the RS-232 channel when entering the diagnostic mode via the keyboard, preventing unnecessary clutter of diagnostic data.