Teledyne Routine Maintenance, Model 300 maintenance schedule, 11.2Replacement of sample filter

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 107
Image 107
11.0 ROUTINE MAINTENANCE

11.0 ROUTINE MAINTENANCE

NOTE

THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!

11.1 Model 300 maintenance schedule

The following are the recommended periodic maintenance items for the Teledyne API Model 300 CO Analyzer:

Date Instrument Was Recieved: ___________________

 

 

 

J

 

 

F

 

 

M

 

 

A

 

 

M

 

 

J

 

 

J

 

 

A

 

 

S

 

 

O

 

 

N

 

 

D

 

 

 

 

ITEM

 

A

 

 

E

 

 

A

 

 

P

 

 

A

 

 

U

 

 

U

 

 

U

 

 

E

 

 

C

 

 

O

 

 

E

 

RECOMMENDED

 

 

 

 

N

 

 

B

 

 

R

 

 

R

 

 

Y

 

 

N

 

 

L

 

 

G

 

 

P

 

 

T

 

 

V

 

 

C

 

ACTION

 

 

Particulate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Repalce weekly or as

 

 

filter element

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

needed.

 

 

Pump

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Replace every 6

 

 

diaphragms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

months.

 

 

Sample flow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Check for proper flow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(800cc/min ±10%)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

annually.

 

 

Pneumatic

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Examine and clean as

 

 

lines

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

necessary.

 

 

Leak Check

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Leak Check after

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

maintenance and at

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

least annually.

 

 

Clean optical

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

As needed.

 

 

bench

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TABLE 11.1

 

 

 

 

 

 

 

 

 

 

 

11.2Replacement of sample filter

a.Turn off the Analyzer. This will stop the pump and eliminate the possibility of sucking debris into the Analyzer while changing the filter element.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 101

Page 107
Image 107
Teledyne Routine Maintenance, Model 300 maintenance schedule, 11.2Replacement of sample filter, Particulate, needed

FAQ

Who should perform the maintenance outlined in this manual?

The operations outlined in this chapter should be performed by qualified maintenance personnel only.

What is the recommended maintenance schedule for Model 300?

The manual provides the recommended periodic maintenance items for the Teledyne API Model 300 CO Analyzer.

How do I replace the sample filter?

Refer to page 101 of the Teledyne API Model 300 CO Analyzer Instruction Manual for detailed instructions on how to replace the sample filter.