Triggering Events

 

 

Triggering Events

 

 

 

 

 

Name

Description

 

ATIMER

Automatic timer expired

EXITZR

Exit zero calibration

EXITSP

Exit span calibration

EXITMP

Exit multi-point calibration

SLPCHG

Slope changed

REPORT PERIOD

This property specifies the period between reports (DAS entries). Enter the number of days and press ENTR. Then enter the number of hours (the leftmost digits) and minutes and press ENTR.

NUMBER OF RECORDS

This property specifies how many records you want to store in this data channel. Each record consists of a time stamp and the sampled measurements from each parameter. The memory of the M300 has capacity for a total of about 8,000 records (for all data channels). When the specified number of records for a data channel has been reached, the channel will “wrap around”, with the oldest records being overwritten with the newest records.

NOTE

Changing this property requires that any existing data records

be deleted because memory must be re-initialized.

Enter the number of records desired and press ENTR. Only when you press ENTR is the data file actually deleted. If you press EXIT the property will not be modified and data file will not be deleted.

RS-232 REPORT

This property controls printing of reports on the RS-232 channel. If set to ON then every time a new report is stored it’s also printed on the RS-232 channel. If this property is set to OFF, the reports are still stored in the data file, but not printed on the RS-232 channel.

PARAMETERS

This property shows the number of parameters being sampled. Each data channel can sample up to 10 parameters, each with a separate sampling mode.

This parameter also specifies the mode of data collection. When sampling a reading such as a concentration at periodic intervals, it is desirable to use the AVG mode to generate an average of the readings during the interval. But when sampling the concentration when

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 46

Page 52
Image 52
Teledyne 300 instruction manual Triggering Events, Name, Description, be deleted because memory must be re-initialized

FAQ

What is the function of the ATIMER property?

The ATIMER property indicates when the automatic timer has expired.

How many records can be stored in the data channel?

The M300 has a capacity for a total of about 8,000 records for all data channels.

How can I control the printing of reports on the RS-232 channel?

You can control the printing of reports on the RS-232 channel by setting the RS-232 REPORT property to ON or OFF.