Teledyne 300 instruction manual Repeatedly Pressing Entr To Get The

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 13
Image 13
REPEATEDLY PRESSING ENTR TO GET THE

NOTE

REPEATEDLY PRESSING ENTR TO GET THE

INSTRUMENT TO DISPLAY THE CORRECT SPAN/ZERO VALVE DOES NOT IMPROVE THE ACCURACY OF THE CALIBRATION, NOR DOES IT SPEED UP STABILIZATION. THE REASON FOR WAITING 5-10 MIN. IS THAT IT TAKES THE INSTRUMENT A PERIOD OF TIME TO ESTABLISH AN ACCURATE AVERAGE FOR THE SPAN/ZERO READING.

11.Adjust the Analyzer span point.

a.Input span gas of approximately 80% of the related full scale range concentration from a known source through the sample port of the Analyzer.

b.Go to manual calibration by pressing the CAL button while in the sample mode.

c.Enter the operator password (512).

d.The menu should now show SPAN, CONC, and EXIT. Enter the concentration of the CO calibration gas by pressing CONC and entering the value from the keyboard.

e.Wait 10 minutes for a stable reading to be attained and then press SPAN, followed by ENTR. If the SPAN button is not displayed, this means that the Analyzer is too far out of adjustment to do a reliable calibration and thus it is not permitted (see Section 3.1 for information on the calibration limits). The operator can exit the manual calibration procedure only by pressing the EXIT button.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 7

Page 13
Image 13
Teledyne 300 instruction manual Repeatedly Pressing Entr To Get The

FAQ

Why does repeatedly pressing ENTR not improve calibration accuracy?

Pressing ENTR multiple times does not improve calibration accuracy or speed up stabilization. It is important to wait 5-10 minutes for the instrument to establish an accurate average for the span/zero reading.

How to adjust the Analyzer span point?

Follow the specific instructions for inputting the span gas, entering manual calibration, entering the operator password, entering the CO calibration gas concentration, and waiting 10 minutes for a stable reading before pressing SPAN.

How to exit the manual calibration procedure?

You can only exit the manual calibration procedure by pressing the EXIT button.