Teledyne 300 instruction manual Signal I/O, Diagnostic Tests, Test #, Nominal Value, Name

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 60
Image 60

The TEST button is used to scroll through the test measurements until the one of interest is displayed. To turn the test on press the OFF/ON button. Viewing test measurements in the diagnostic mode is especially useful for viewing the results of a diagnostic test.

The PREV button goes to the previous diagnostic test. When pressed, the CPU turns the current diagnostic test OFF if it is ON. The NEXT button goes to the next diagnostic test. When pressed, the CPU turns the current diagnostic test OFF if it is ON. The EXIT button exits the diagnostic mode and turns all the diagnostic tests OFF. This ensures that a diagnostic test is not accidentally left ON. A message is also sent to the RS-232 channel to indicate that the diagnostic mode has been exited.

DIAGNOSTIC TESTS

Test #

Name

Signal

Nominal Value

1

NONE

No output

0 mV

2

CO MEASURE

CO detector measure value

4500 mV

3

CO REFERENCE

CO detector reference

4200 mV

 

 

value

 

4

SAMP PRESS

Sample pressure

125 mV

5

SAMP FLOW

Sample flow

3000 mV

6

SAMP TEMP

Sample temperature

3000 mV

7

BENCH TEMP

Optical bench temperature

3900 mV

8

WHEEL TEMP

Outputs filter wheel temp.

3900 mV

9

CHASSIS TEMP

Outputs Chassis temp

2740 mV

10

DCPS VOLT

Outputs DC power

2500 mV

11

DAS AVERAGE

Current DAS average

Any

TABLE 5.1

5.2.1 Signal I/O

The signal I/O diagnostic mode gives the user access to the digital and analog inputs and outputs on the V/F board. The digital outputs can be controlled through the keyboard. Any signals manually changed through the signal I/O menu will remain in effect until you leave the signal I/O menu. At that time the analyzer will regain control of these signals. To enter the signal I/O test mode, press SETUP-MORE-DIAG-ENTR. When the diagnostic mode is entered, a message is sent to the RS-232 channel indicating entry into the diagnostic mode. Use the PREV and NEXT buttons to scroll through the signals. Edit buttons will appear for the signals that can be controlled by the user. Press JUMP to skip to a specific I/O Signal.

Table 5.2 below lists the I/O signals available for the M300:

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 54

Page 60
Image 60
Teledyne 300 instruction manual Signal I/O, Diagnostic Tests, Test #, Nominal Value, Name

FAQ

What is the TEST button used for?

The TEST button is used to scroll through the test measurements until the one of interest is displayed.

How do I exit the diagnostic mode?

Press the EXIT button to exit the diagnostic mode and turn all the diagnostic tests OFF.

What signals can be controlled in the signal I/O mode?

The signal I/O mode gives the user access to digital and analog inputs and outputs on the V/F board.