Teledyne 300 instruction manual Check That Analyzer Is Set Up For Proper, Voltage And Frequency

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 12
Image 12
CHECK THAT ANALYZER IS SET UP FOR PROPER

CAUTION

CHECK THAT ANALYZER IS SET UP FOR PROPER

VOLTAGE AND FREQUENCY.

CAUTION

POWER PLUG MUST HAVE GROUND LUG.

7.Turn on the M300 by switching the switch on the lower right corner of the front panel (See Figure 2.1). The front panel display should light with a sequence of messages, -API - M300 - software version number, then a normal display as shown in Figure 2.2.

8.Allow about 60 minutes for the temperatures to come up to their respective setpoints then press the left most button on the front keyboard to scroll through the TEST values. Compare these values to those noted during the final factory checkout listed in Table 1.1. The values observed should closely match the Table 1.1 values.

9.Select the range on which the analyzer will be calibrated.

a.From the SAMPLE menu press SETUP to enter the SETUP menu. (See Figure 2.2 for appearance of front panel.)

b.Enter the PASSWORD (818).

c.Press RNGE (RANGE).

d.Press SET.

e.Enter the derived full scale range for analog outputs and press ENTR.

f.Press EXIT 2 times to return to the SAMPLE menu.

10.Adjust the analyzer zero point.

a.Input zero air into the sample port.

b.Press CAL from the SAMPLE menu and enter the password (818).

c.Press ZERO.

d.It usually takes about 5 to 10 minutes for the reading to stabilize near zero. After a stable reading has been obtained press ENTR. The display should now read 0.0 ppm carbon monoxide.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 6

Page 12
Image 12
Teledyne 300 Check That Analyzer Is Set Up For Proper, Voltage And Frequency, Power Plug Must Have Ground Lug

FAQ

What should I check before setting up the analyzer?

Before setting up the analyzer, you should check that the analyzer is set up for proper voltage and frequency, and ensure that the power plug has a ground lug.

How do I turn on the M300 analyzer?

To turn on the M300 analyzer, switch the switch on the lower right corner of the front panel. The front panel display should light with a sequence of messages and software version number.

How do I adjust the analyzer zero point?

To adjust the analyzer zero point, input zero air into the sample port, press CAL from the SAMPLE menu, enter the password, press ZERO, and wait for the reading to stabilize near zero.