SYSTEM MODES

 

 

 

SYSTEM MODES

 

 

 

 

 

 

 

Mode

 

Meaning

 

 

SAMPLE x (1)

 

Sampling normally

 

 

SAMPLE x (1)

 

Flashing indicates adaptive filter is on

 

 

ZERO CAL x (2)

 

Doing a zero check or adjust

 

 

SPAN CAL x (2)

 

Doing a span check or adjust

 

 

MP CAL

 

Doing a multi-point calibration

 

 

SETUP xxx (3)

 

Configuring analyzer (sampling continues)

 

 

DIAG DAS (4)

 

Data Acquisition configuration

 

 

DIAG I/O (4)

 

Test digital I/O signals

 

 

DIAG AOUT (4)

 

Test analog output channels

 

 

DIAG D/A (4)

 

Configure and Calibrate Digital to Analog converters

 

 

DIAG TCHN (4)

 

Configure Test Channel output

 

 

 

 

(1) x = A (auto)

 

 

(2)

x = M (manual), A (auto), R (remote)

 

(3) xxx = software revision (e.g. A.9)

(4) diagnostic test modes

TABLE 2.2

The message field shows test measurements or warning messages. Tables 2.3 and 2.4 summarize the test measurements and warning messages and their meanings. Refer to Sections 4.0 and 5.0 for detailed information on viewing test measurements and warning messages and clearing warnings.

 

TEST MEASUREMENTS

 

 

 

Test Message

 

Meaning

TIME=xx:xx:xx

 

Current time-of-day (HH:MM:SS)

RANGE=xxxx PPM

 

Analog output full-scale range

STABIL=x.xxx PPM

 

Standard Deviation of CO readings

CO MEAS=xxxxx MV

 

Current V/F measure channel (mV)

CO REF=xxxxx MV

 

Current V/F reference channel (mV)

MR RATIO=X.XXX

 

Ratio of the Reference and Measure values

PRES=xxx IN-HG-A

 

Absolute sample pressure - inches Hg

SAMPLE FL=xxx CC/M

 

Sample flow through Analyzer (cc/min)

SAMPLE TEMP=xxx C

 

Temperature in the absorption cell

BENCH TMP=xxx C

 

Optical Bench Temperature

WHEEL TMP=xxx C

 

Filter Wheel Temperature

BOX TEMP=xxx C

 

Internal box temperature (degrees C)

DCPS=xxxxxx MV

 

DC power supply (mV)

SLOPE=x.xxx

 

Internal formula - Slope

OFFSET=xx.x MV

 

Internal formula - Offset

 

 

TABLE 2.3

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 26

Page 32
Image 32
Teledyne 300 instruction manual System Modes, Meaning, Test Measurements

FAQ

What are the different system modes?

The system modes include Sample x 1, Span Cal x 2, Setup xxx 3, and more. Each mode has a specific meaning and purpose.

How can I check test measurements?

You can check test measurements such as current time-of-day, CO readings, temperature, and more using the Teledyne API Model 300 CO Analyzer.

Where can I find detailed information on warning messages?

Detailed information on warning messages and how to clear them can be found in Sections 4.0 and 5.0 of the instruction manual.