Teledyne 300 instruction manual Contro, Signal, Description

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 61
Image 61

#

Signal

Contro

Description

 

 

 

l

 

 

0

DISP_BROWNOUT

NO

Display brownout is used to keep the

 

 

 

display from getting corrupted during

 

 

 

low line voltage conditions. Circuitry

 

 

 

on the Power Supply board (00015)

 

 

 

senses low line voltage and sets this

 

 

 

bit. The CPU reads this and generates

 

 

 

the BROWNOUT_RST signal

 

 

 

described below.

 

1

EXT_ZERO_CAL

NO

Shows state of status input bit to cause

 

 

 

the M300 to enter Zero Calibration

 

 

 

mode. Use to check external contact

 

 

 

closure circuitry.

 

2

EXT_SPAN_CAL

NO

Shows state of status input bit to cause

 

 

 

the M300 to enter the Span Calibration

 

 

 

mode. Use to check external contact

 

 

 

closure circuitry.

 

3

SYNC_OK

NO

Indicates that demodulation circuitry

 

 

 

on the Sync/Demod Board (00798) is

 

 

 

able to lock-in on the detector signal.

4

SPAN_VALVE

YES

Switches the Zero/Span valve.

Use

 

 

 

this bit to test the valve function.

 

5

CAL_VALVE

YES

Switches the Sample/Cal valve.

Use

 

 

 

this bit to test the valve function.

 

6

BENCH_HTR

YES

Shows the status of the optical bench

 

 

 

heater. This has the same function as

 

 

 

the LED in the power supply module.

7

WHEEL_HTR

YES

Shows the status of the filter wheel

 

 

 

heater. This has the same function as

 

 

 

the LED in the power supply module.

8

DARK_CAL

YES

Turns off the detector input to the

 

 

 

Sync/Demod board for electronics

 

 

 

calibration.

 

9

ELEC_TEST

YES

Activates the Electric Test diagnostic

 

 

 

circuitry.

 

10

BROWNOUT_RST

YES

Resets the DISP_BROWNOUT

 

 

 

circuitry described above.

 

11

ST_ZERO_CAL

YES

Status Bit - Zero Calibration mode

 

 

 

 

Logic high = M300 in Zero cal mode

 

 

 

Logic low = Not in Zero cal mode

 

12

ST_SPAN_CAL

YES

Status Bit - Span Calibration mode

 

 

 

 

Logic high = M300 in Span cal mode

 

 

 

Logic low = Not in Span cal mode

 

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 55

Page 61
Image 61
Teledyne 300 instruction manual Contro, Signal, Description

FAQ

What is the purpose of DISP_BROWNOUT?

Display brownout is used to prevent display corruption during low line voltage conditions. It is controlled by the Power Supply board 00015.

How do I activate the Electric Test diagnostic circuitry?

You can activate the Electric Test diagnostic circuitry by using the ELEC_TEST signal on the alarm panel.

How can I reset DISP_BROWNOUT circuitry?

To reset the DISP_BROWNOUT circuitry, use the BROWNOUT_RST signal on the alarm panel.