Teledyne 300 Data Parameters, Name, Units, None, Comeas, CO Measure signal, Coref, SLOPE1, SLOPE2

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 54
Image 54
Data Parameters

Data Parameters

Name

Description

Units

NONE

 

 

COMEAS

CO Measure signal

mV

COREF

CO Reference signal

mV

SLOPE1

Slope of Range #1

 

SLOPE2

Slope of Range #2

 

OFSET1

Offset of Range #1

mV

OFSET2

Offset of Range #2

mV

ZSCNC1

Range1 zero/span measured

PPM

 

concentration

 

ZSCNC2

Range2 zero/span measured

PPM

 

concentration

 

CONC1

Concentration of Range #1

PPM

CONC2

Concentration of Range #2

PPM

STABIL

Stability

PPM

BENCHTMP

Optical bench temperature

oC

WHEELTMP

Filter wheel temperature

oC

SMPFLW

Sample flowrate

cc/min

SMPPRS

Sample pressure

in-Hg

BOXTMP

Box temperature

oC

DCPS

DC power supply composite voltage

mV

Example:

Sample the CO concentration (Range #1) once per minute (default) and issue a report once per 5 minutes, and print the report on the RS-232 channel.

1.Press SETUP-DAS-EDIT-ENTR-INSto create a new data channel and begin editing its properties.

2.Edit the NAME property and give it any name you want, but do not set the blank name (all dashes) or duplicate names because they won’t be accessible from the RS- 232 interface.

3.For each of the remaining properties, press the SET> to view the current setting, and then press EDIT to change the property if necessary. The table below lists the proper settings for all of the properties for this example:

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 48

Page 54
Image 54
Teledyne 300 Data Parameters, Name, Units, None, Comeas, CO Measure signal, Coref, CO Reference signal, SLOPE1, SLOPE2

FAQ

What are the data parameters for Teledyne 300 Carbon Monoxide Alarm?

The data parameters include CO Measure signal, CO Reference signal, range slopes, offsets, zero/span measurements, concentration values, stability, and temperature readings.

How often should the CO concentration be sampled?

The CO concentration Range #1 is sampled once per minute by default, and a report is issued once per 5 minutes, which can be printed on the RS-232 channel.

How can I create a new data channel for the Teledyne 300 CO Analyzer?

To create a new data channel, press SETUP-DAS-EDIT-ENTR-INS and begin editing its properties. Edit the NAME property and avoid setting a blank name, all dashes, or duplicate names as they won't be accessible from the RS-232 interface.