Teledyne 300 instruction manual Adjust R7 for a CO MEAS reading of 4200 ±200 mV

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 104
Image 104

5.Confirm the wave from of the optical signal by attaching an oscilloscope to the Sync Demodulator board a U2 Pin 7. The oscilloscope trace should appear like those shown in Figure 10.2. In particular the wave form should be symmetrical and should have distinct flat regions at the top and bottom of the pulses.

6.If unable to achieve 4200 mV on CO MEAS, then do the following:

7.Adjust source for maximum signal strength. Typically source has no effect on wave shape. (Not to exceed 27V peak to peak.)

8.Adjust input mirror as needed to create a wave shape with distinct flattening of the peaks and symmetrical “Knees” at the peaks. A smaller waveform with nice symmetry is preferable to a larger one which is asymmetrical.

9.With zero air in, verify an MR RATIO of 1.18 to 1.22 is desired. If the analyzer shows a significantly different value, slightly adjust the input mirror.

10.Adjust R7 for a CO MEAS reading of 4200 ±200 mV.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 98

Page 104
Image 104
Teledyne 300 instruction manual Adjust R7 for a CO MEAS reading of 4200 ±200 mV

FAQ

What should the oscilloscope trace look like?

The waveform should be symmetrical with distinct flat regions at the top and bottom of the pulses.

What to do if unable to achieve 4200 mV on CO MEAS?

Adjust the source for maximum signal strength, typically source has no effect on wave shape. Not to exceed 27V peak to peak.

What is the desired MR ratio with zero air in?

A MR ratio of 1.18 to 1.22 is desired. If the analyzer shows a significantly different value, slightly adjust the input mirror.