WARNING MESSAGES

WARNING MESSAGES

 

Warning Message

 

 

Meaning

 

 

 

 

 

 

SYSTEM RESET

 

 

Issued whenever Analyzer is powered on

 

 

RAM INITIALIZED

 

RAM was erased (incl. DAS reports)

 

SOURCE WARNING

 

IR source < 2500 OR >= 5000 mV

 

BENCH HEAT SHUTDOWN

 

Optical bench temp control not working

 

SAMPLE FLOW WARNING

 

Sample flow < 500 cc/m or > 1000 cc/m

 

SAMPLE PRESSURE WARN

 

Sample pressure < 15 or > 35 In-Hg-A

 

SAMPLE TEMP WARNING

 

Sample temperature < 10°C or > 50°C

 

BOX TEMP WARNING

 

Box temp. < 12°C or > 48°C

 

BENCH TEMP WARN

 

Optical Bench < 43°C or > 53°C

 

WHEEL TEMP WARN

 

Filter Wheel < 63°C or > 73°C

 

CANNOT DYN ZERO

 

CO offset < -1500 or > +1500 mV

 

CANNOT DYN SPAN

 

CO slope < 0.5 or > 2.0

 

V/F NOT INSTALLED

 

A/D - I/O card not installed or bad

 

SYNC ERROR

 

No modulation on detector output

 

 

 

 

TABLE 2.4

The menu field changes depending on the mode of the Analyzer and the buttons that have been pressed. It indicates the current function of each of the 8 push buttons below the display. See Section 2.2.2 for information on using the push buttons.

2.2.2 Programmable push buttons

The 8 push buttons below the display are programmable by the CPU in that their functions change depending on the mode of the Analyzer or the operations being performed. The legend above a button identifies its current function. If there is no legend above a button, it has no function and will be ignored if pressed.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 27

Page 33
Image 33
Teledyne 300 instruction manual Programmable push buttons, Warning Messages, Meaning

FAQ

What are the warning messages for Teledyne API Model 300 CO Analyzer?

The warning messages include SYSTEM RESET, RAM INITIALIZED, SOURCE WARNING, BENCH HEAT SHUTDOWN, SAMPLE FLOW WARNING, SAMPLE PRESSURE WARNING, SAMPLE TEMP WARNING, BOX TEMP WARNING, BENCH TEMP WARNING, WHEEL TEMP WARNING, CANNOT DYN ZERO, CANNOT DYN SPAN, V/F NOT INSTALLED, and SYNC ERROR.

How can I change the functions of the programmable push buttons on Teledyne API Model 300 CO Analyzer?

The functions of the push buttons change depending on the mode of the Analyzer or the operations being performed. See Section 2.2.2 of the manual for information on using the push buttons.

Where can I find more information about Teledyne API Model 300 CO Analyzer?

You can find more information about Teledyne API Model 300 CO Analyzer in the official instruction manual, particularly in Section 2.2.2 for programmable push buttons and their functions.