1.5 Installation and overview

1.5 Installation and overview

The Model 300 is shipped with the following standard equipment:

1.Power cord.

2.Instruction manual.

CAUTION

To avoid personal injury, always use two persons to lift and

carry the Model 300.

Upon receiving the Model 300 please do the following:

1.Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)

2.When installing the Model 300, allow a minimum of 4 inches (100mm) of clearance at the back of the instrument and 1 inch (25mm) of clearance on each side for proper ventilation. Also, be sure that the clearance below the chassis is unobstructed by at least the height of the instrument feet.

3.Connect sample inlet line to the sample port on rear panel.

NOTE

SEE FIGURE 1.4 FOR REAR PANEL PNEUMATIC

CONNECTIONS. SAMPLE GAS SHOULD ONLY COME

INTO CONTACT WITH PTFE, GLASS OR STAINLESS

STEEL. LEAK CHECK ALL FITTINGS WITH SOAP

SOLUTION.

MAXIMUM PRESSURE FOR LEAK CHECK IS 5 PSI.

CAUTION

Connect the exhaust fitting on the rear panel (See Fig. 1.4) to a

suitable vent outside the analyzer area.

4.Connect IZS inlet input (if installed) to a clean, dry air supply.

5.Connect a recording device to the terminal strip connections on the rear panel (See Figure 1.2).

6.Connect the power cord to an appropriate power outlet (see the serial number tag for correct voltage and frequency).

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 5

Page 11
Image 11
Teledyne 300 Installation and overview, SEE .4 FOR REAR PANEL PNEUMATIC, Connections. Sample Gas Should Only Come

FAQ

What standard equipment is shipped with the Model 300?

The Model 300 is shipped with a power cord and instruction manual.

How much clearance is required for proper ventilation during installation?

A minimum of 4 inches (100mm) of clearance at the back of the instrument and 1 inch (25mm) of clearance on each side is required. Additionally, the clearance below the chassis should be unobstructed by at least the height of the instrument feet.

What should be used for leak checking the fittings on the rear panel?

A soap solution should be used for leak checking all fittings, with a maximum pressure of 5 PSI.