DESCRIPTION

 

Property

 

Setting

 

 

DESCRIPTION

 

 

 

 

 

 

 

NAME

 

“CO5MIN”

 

 

 

 

 

EVENT

ATIMER

 

Sample when automatic

 

 

 

 

 

 

timer is expired

 

PARAMETERS

CONC1,AVG

SO2 Range #1 concentration

 

REPORT PERIOD

000:00:05 (i.e. 5 minutes)

Average of 5 minutes*

 

 

 

 

 

 

reported every 5 minutes

 

NUMBER OF

4032 (i.e. two weeks)

Stores 4032 of Report

 

RECORDS

 

 

 

Period (5 minutes average)

 

RS-232 REPORT

ON

 

RS-232 is enabled

 

CHANNEL

ON

 

Data channel is enabled

 

ENABLED

 

 

 

 

 

 

CAL. HOLD OFF

ON

 

Hold off is enabled

*Sample measurement period of filtered CO concentration is 1 minute by default.

4.11.3 RS-232 reporting

Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel. For all default data channels, RS-232 reporting is initially set to “OFF.” If this property is turned on, the Data Channel will issue a report with a time and date stamp to the RS-232 port every time a data point is logged. The report format is shown below:

D 31:10:06 0412 CONC : AVG COCNC1=6.8 PPB

The report consists of the letter “D” follow by a time/date stamp(“31:10:06”) followed by the instrument ID number(“0412”). Next is the Data Channel name(“CONC”) and the sampling mode (“AVG” indicates that the data point is an average of more than one sample as opposed to a instantaneous reading, “INST”). Finally, the name of the parameter and its value(“COCNC1=6.8 PPB”) are printed. For Data Channels that sample more than one parameter, such as PNUMTC and CALDAT, each parameter is printed on a separate line.

To enable RS-232 reporting for a specific Data Channel:

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 49

Page 55
Image 55
Teledyne 300 instruction manual 4.11.3 RS-232reporting, Description, Property, Setting

FAQ

What is the default reporting period for RS-232 reporting?

The default reporting period is 5 minutes.

How can I enable RS-232 reporting for a specific Data Channel?

To enable RS-232 reporting for a specific Data Channel, refer to page 49 of the Teledyne API Model 300 CO Analyzer Instruction Manual.

What is the format of the RS-232 report?

The report consists of a time/date stamp, instrument ID number, Data Channel name, sampling mode, and the parameter value.