5.0 DIAGNOSTICS

5.0 DIAGNOSTICS

The Teledyne API carbon monoxide analyzer contains two levels of diagnostics: test measurements which can be viewed at all times (except when in setup) by pressing TEST, and lower level diagnostic operations which can only be performed by pressing

SETUP-DIAG.

5.1 Test measurements

As stated, test measurements can be viewed at any time except when in setup. To view a different test measurement, simply press the TEST button. Table 2.3 lists the test measurements which are available. Viewing these test measurements does not interfere with the operation of the Model 300 or the carbon monoxide reading in any way, so they may be viewed freely.

NOTE

IF THE VALUE OF ANY TEST FUNCTION IS DISPLAYED

AS “XXXX”, THIS INDICATES THAT THE READING IS OFF

SCALE OR OTHERWISE NON-VALID.

Additionally, the values of most TEST functions can output as an analog voltage at the instrument's rear panel (see Figure 1.2). The TEST function to be output is selected by pressing SETUP-MORE-DIAG. Press NEXT until TEST CHANNEL OUTPUT appears. Press ENTR. Select test channel function and press ENTR. Table 5.1 below lists the Test functions available for analog output.

In addition to outputting a value to the analog output channel, these tests activate a new test measurement which displays the analog voltage reading on the front panel as:

"TEST=XXXXX.X MV".

When you exit the diagnostics, this test measurement is removed.

5.2 Diagnostic tests

The diagnostic tests are used to help diagnose a problem in the Analyzer and should only be used by skilled maintenance people since they can potentially interfere with the carbon monoxide reading. Table 5.1 below lists the low level diagnostic tests which are available. To get into the diagnostic test mode, press SETUP-DIAG. When the diagnostic mode is entered, a message is sent to the RS-232 channel indicating entry into the diagnostic mode. The buttons which are available to the operator are described below.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 53

Page 59
Image 59
Teledyne 300 instruction manual Diagnostics, Test measurements, Diagnostic tests, Setup-Diag

FAQ

Can I view test measurements at any time?

Yes, test measurements can be viewed at any time except when in setup.

How do I select a test channel function for analog output?

Press SETUP-MORE-DIAG, then press NEXT until TEST CHANNEL OUTPUT appears. Select the test channel function and press ENTR.

Who should use the diagnostic tests?

The diagnostic tests should only be used by skilled maintenance personnel as they can potentially interfere with the carbon monoxide reading.