Teledyne 300 instruction manual Test measurements, Test Measurement Request Commands

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 73
Image 73
7.5 Test measurements

7.5 Test measurements

All the test measurements which can be displayed by pressing the TEST button are also available to the host computer via the RS-232 interface. The host computer should issue a request for a test measurement, and then the CPU will send the current value of the test measurement to the RS-232 output. The format of the test measurement message is:

"T DDD:HH:MM IIII TEST MEASUREMENT<CRLF>"

For example, the format of the DC Power Supply output in millivolts would be:

"T 194:11:29 0000 DCPS= 2500 MV<CRLF>"

To request a test measurement, the host must issue a command of the form:

"T MEASUREMENT<CRLF>"

For a summary of all test functions issue the command "T LIST". The table below lists the commands and the corresponding test measurements which will be returned.

TEST MEASUREMENT REQUEST COMMANDS

Command

 

Test measurement

"?<CRLF>"

 

RS-232 HELP screen

"T LIST<CRLF>"

 

Summary of all TEST's

"T CO<CRLF>"

 

Current CO reading

"T COMEAS<CRLF>"

 

Current CO MEAS mV

"T COREF<CRLF>"

 

Current CO REF mV

"T MRRATIO<CRLF>"

 

Current MR RATIO

"T SPRESS<CRLF>"

 

Sample pressure

"T SFLOW<CRLF>"

 

Sample flow rate

"T STEMP<CRLF>"

 

Sample temperature

"T BNCHTEMP<CRLF>"

 

Optical Bench temperature

"T WHEELTEMP<CRLF>"

 

Filter Wheel temperature

"T BOXTEMP<CRLF>"

 

Internal box temperature

"T DCPS<CRLF>"

 

DC power supply output

"T COSLOPE<CRLF>"

 

Slope value

"T COFFSET<CRLF>"

 

Offset value

"T CLKTIME<CRLF>"

 

Current time-of-day

 

TABLE 7.7

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 67

Page 73
Image 73
Teledyne 300 instruction manual Test measurements, Test Measurement Request Commands

FAQ

What test measurements are available for the Teledyne API Model 300 CO Analyzer?

The test measurements available include current CO reading, CO MEAS mV, CO REF mV, MR RATIO, sample pressure, sample flow rate, sample temperature, optical bench temperature, filter wheel temperature, internal box temperature, DC power supply output, slope value, offset value, and current time-of-day.

How can I request a test measurement from the Teledyne API Model 300 CO Analyzer?

You can request a test measurement by issuing a command of the form 'T MEASUREMENT' to the host computer via the RS-232 interface.

What is the format of the test measurement message for the Teledyne API Model 300 CO Analyzer?

The format of the test measurement message is 'T DDD:HH:MM IIII TEST MEASUREMENT'. For example, the format of the DC Power Supply output in millivolts would be 'T 194:11:29 0000 DCPS= 2500 MV'.