Leica 1000Z01 user manual TPS-System 1000 Programs-2.3.1en

Models: 1000Z01

1 347
Download 347 pages 59.8 Kb
Page 57
Image 57

Displays date and version of the running application.

Set the values to default. Default values are displayed in dialog on page 55.

Exit the program.

Store the current configuration and proceed to the dialog "MEASURE MODE".

Dual-face Measurement In the dual-face mode, the program will prompt for measurements in both faces. When both measurements are taken, the program will check the difference between the two. If the difference in angle is within 27' (0.5 gon)) and the difference of two measured distances is within 0.5 m (1.64 ft), the observations will be averaged. These tolerances are used to avoid errors in target identification. If exceeded an error message will be displayed.

56

TPS-System 1000 Programs-2.3.1en

© Leica

Page 57
Image 57
Leica 1000Z01 user manual TPS-System 1000 Programs-2.3.1en