Configuration

Configuration Editor

Start the "Configuration Editor" from the "TARGET-

POINT" dialog.

OH

ORINT\ CONFIGURATION

 

 

14:03

 

 

 

 

 

 

 

 

 

 

Hz Ori Acc

:

0°00'32"

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ht Acc TP

:

0.0250

 

m

 

 

 

 

 

Posn Acc TP :

0.0250

 

m

 

 

 

 

 

Two faces

:

 

 

 

NO

 

 

 

 

 

 

 

User Displ

:

 

 

 

NO

 

 

 

 

 

 

 

Log File

:

 

 

OFF

 

 

 

EDIT

INFO

 

 

 

 

 

DFLT

 

 

 

Log FlName

:

ORIENT.LOG

 

 

 

 

 

 

 

MC

HELP

F1

F2

F3

F4

F5

F6

The "Configuration Editor "sets parameters for further program operations:

Hz Ori Acc : Limit for the standard deviation of the orientation. The orientation is regarded as "error free", if the computed standard deviation of the orientation is within twice the entered value.

Ht Acc TP : Height accuracy of the target points. The entered value, is used as an "a priori" accuracy in the calculation. The height is regarded as "error free", if the computed standard deviation is within twice the entered value.