Basic Operation

Model GFC7001E Carbon Dioxide Analyzer

6.6.3.1.RNGE MODE SNGL: Configuring the GFC 7001E/EM Analyzer for SINGLE Range Mode

NOTE

This is the default reporting range mode for the analyzer.

When the single range mode is selected (SNGL), all analog CO concentration outputs (A1 and A2) are slaved together and set to the same reporting range limits (e.g. 500.0 ppb). The span limit of this reporting range can be set to any value within the physical range of the analyzer.

Although both outputs share the same concentration reporting range, the electronic signal ranges of the analog outputs may still be configured for different values (e.g. 0-5 VDC, 0-10 VDC, etc; see Section 7.4.2)

To select SNGL range mode and to set the upper limit of the range, press:

Teledyne Analytical Instruments

100

Page 100
Image 100
Teledyne GFC 7001EM operation manual This is the default reporting range mode for the analyzer