Model GFC7001E Carbon Dioxide Analyzer

An alternative method for measuring the output of the Current Loop converter is to connect a 250 ohm 1% resistor across the current loop output in lieu of the current meter (see Figure 3-7 for pin assignments and diagram of the analog output connector). This allows the use of a voltmeter connected across the resistor to measure converter output as VDC or mVDC.

V

+DC Gnd

Figure 7-7: Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels

In this case, follow the procedure above but adjust the output for the following values:

Table 7-9: Current Loop Output Check

% FS

Voltage across

Voltage across

Resistor for 2-20 mA

Resistor for 4-20 mA

 

0

500 mVDC

1000 mVDC

100

5000 mVDC

5000 mVDC

Teledyne Analytical Instruments

141

Page 141
Image 141
Teledyne GFC 7001EM Voltage across Resistor for 2-20 mA Resistor for 4-20 mA, Mvdc, Current Loop Output Check