Advanced Features Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 144

7.4.6. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4

The test functions available to be reported are listed in Table 7-10:

Table 7-10: Test Channels Functions available on the GFC 7001E/EM’s Analog Output

TEST CHANNEL DESCRIPTION ZERO FULL SCALE *

NONE TEST CHANNEL IS TURNED OFF.

CO MEASURE
The demodulated, peak IR detector output
during the measure portion of the GFC Wheel
cycle.
0 mV 5000 mV
CO REFERENCE
The demodulated, peak IR detector output
during the reference portion of the GFC
Wheel cycle.
0 mV 5000 mV
SAMPLE PRESS
The absolute pressure of the Sample gas as
measured by a pressure sensor located inside
the sample chamber.
0" Hg 40 "Hg
SAMPLE FLOW Sample mass flow rate as measured by the
flow rate sensor in the sample gas stream. 0 cm3/m 1000 cm
3/m
SAMPLE TEMP The temperature of the gas inside the sample
chamber. 0C 70C
BENCH TEMP Optical bench temperature. 0C 70C
WHEEL TEMP GFC Wheel temperature. 0C 70C
O2 CELL TEMP The current temperature of the O2 sensor
measurement cell. n 70C
CHASSIS TEMP The temperature inside the analyzer chassis. 0C 70C
PHT DRIVE
The drive voltage being supplied to the
thermoelectric coolers of the IR photo-
detector by the Sync/Demod Board.
0 mV 5000 mV
* Maximum test signal value at full scale of test channel output.

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds

TEST to the list of test functions viewable via the front panel display.