Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

7.4.6.SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4

The test functions available to be reported are listed in Table 7-10:

Table 7-10: Test Channels Functions available on the GFC 7001E/EM’s Analog Output

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE *

 

 

 

 

NONE

TEST CHANNEL IS TURNED OFF.

 

 

CO MEASURE

The demodulated, peak IR detector output

 

 

during the measure portion of the GFC Wheel

0 mV

5000 mV

 

cycle.

 

 

CO REFERENCE

The demodulated, peak IR detector output

 

 

during the reference portion of the GFC

0 mV

5000 mV

 

Wheel cycle.

 

 

SAMPLE PRESS

The absolute pressure of the Sample gas as

 

 

measured by a pressure sensor located inside

0" Hg

40 "Hg

 

the sample chamber.

 

 

SAMPLE FLOW

Sample mass flow rate as measured by the

0 cm3/m

1000 cm 3/m

 

flow rate sensor in the sample gas stream.

 

 

SAMPLE TEMP

The temperature of the gas inside the sample

0C

70C

chamber.

 

 

 

BENCH TEMP

Optical bench temperature.

0C

70C

 

 

 

 

WHEEL TEMP

GFC Wheel temperature.

0C

70C

O2 CELL TEMP

The current temperature of the O2 sensor

n

70C

measurement cell.

 

 

 

CHASSIS TEMP

The temperature inside the analyzer chassis.

0C

70C

PHT DRIVE

The drive voltage being supplied to the

 

 

thermoelectric coolers of the IR photo-

0 mV

5000 mV

 

detector by the Sync/Demod Board.

 

 

* Maximum test signal value at full scale of test channel output.

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds TEST to the list of test functions viewable via the front panel display.