FAQs

Model GFC7001E Carbon Dioxide Analyzer

Q:How do I measure the sample flow?

A:Sample flow is measured by attaching a calibrated rotameter, wet test meter, or other flow-measuring device to the sample inlet port when the instrument is operating. The sample flow should be 800 cm3/min 10%. See Section 12.3.4.

Q:How long does the IR source last? A: Typical lifetime is about 2-3 years.

Q:Can I automate the calibration of my analyzer?

A:Any analyzer with zero/span valve or IZS option can be automatically calibrated using the instrument’s AutoCal feature. The setup of this option is located in Section 9.4.

Q:Can I use the IZS option to calibrate the analyzer?

A:Yes. However, whereas this may be acceptable for basic calibration checks, the IZS option is not permitted as a calibration source in applications following US EPA protocols.

To achieve highest accuracy, it is recommended to use cylinders of calibrated span gases in combination with a zero air source.

Q:My analyzer has the optional, user-configurable analog output channels. How do I program and use them? A: Instructions for this can be found in Appendix E .

Q:What is the averaging time for an GFC 7001E/EM?

A:The default averaging time, optimized for ambient pollution monitoring, is 150 seconds for stable concentrations and 10 seconds for rapidly changing concentrations; See Section 11.5.12 for more information. However, it is adjustable over a range of 0.5 second to 200 seconds (please contact customer service for more information).

4.2.GLOSSARY

 

Term

Description/Definition

 

 

10BaseT

An Ethernet standard that uses twisted (“T”) pairs of copper wires to transmit at

 

 

 

 

10 megabits per second (Mbps).

 

 

100BaseT

Same as 10BaseT except ten times faster (100 Mbps).

 

 

 

 

 

 

APICOM

Name of a remote control program offered by Teledyne to its customers.

 

 

ASSY

Assembly.

 

 

cm3

metric abbreviation for cubic centimeter. Same as the obsolete abbreviation “cc”.

 

 

Chemical formulas that may be included in this document:

 

 

CO2

carbon dioxide

 

 

C3H8

propane

 

 

CH4

methane

 

 

H2O

water vapor

 

 

HC

general abbreviation for hydrocarbon

 

 

HNO3

nitric acid

 

 

H2S

hydrogen sulfide

 

 

NO

nitric oxide

 

 

NO2

nitrogen dioxide

 

 

 

 

 

 

 

 

 

Teledyne Analytical Instruments

58

Page 58
Image 58
Teledyne GFC 7001EM operation manual Glossary, Term Description/Definition