Calibration Procedures

Model GFC7001E Carbon Dioxide Analyzer

9.2.MANUAL CALIBRATION CHECKS AND CALIBRATION OF THE GFC 7001E/EM ANALYZER IN ITS BASE CONFIGURATION

ZERO/SPAN CALIBRATION CHECKS VS. ZERO/SPAN CALIBRATION

Pressing the ENTR key during the following procedure resets the stored values for OFFSET and SLOPE

and alters the instrument’s Calibration.

This should ONLY BE DONE during an actual calibration of the GFC 7001E/EM.

NEVER press the ENTR key if you are only checking calibration.

9.2.1.SETUP FOR BASIC CALIBRATION CHECKS AND CALIBRATION

STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.

Figure 9-1: Pneumatic Connections – Basic Configuration – Using Bottled Span Gas