Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

7.4.USING THE GFC 7001E/EM ANALYZER’S ANALOG OUTPUTS.

The GFC 7001E/EM Analyzer comes equipped with four analog outputs.

The first two outputs (A1 & A2) carry analog signals that represent the currently measured concentration of CO (see Section 6.6.2).

The third output (A3) is only active if the analyzer is equipped with one of the optional 2nd gas sensors (e.g. O2 or CO2).

The fourth output (A4) outputs a signal that can be set to represent the current value of one of several test functions (see Table 7-10).

7.4.1.ACCESSING THE ANALOG OUTPUT SIGNAL CONFIGURATION SUBMENU

The following lists the analog I/O functions that are available in the GFC 7001E/EM Analyzer.

 

 

Table 7-6: DIAG - Analog I/O Functions

 

 

 

 

 

SUB MENU

OUTPUT

FUNCTION

MANUAL

CHANNEL

SECTION

 

 

 

 

Initiates a calibration of the A1, A2, A3 and A4 analog output

 

AOUT

 

channels that determines the slope and offset inherent in the

 

ALL

circuitry of each output.

7.4.3

CALIBRATED

 

These values are stored and applied to the output signals by

 

 

 

 

 

 

the CPU automatically.

 

 

 

 

 

 

 

Sets the basic electronic configuration of the A1 output (CO

 

 

 

Concentration).

 

 

 

There are four options:

 

 

 

RANGE1: Selects the signal type (voltage or current loop)

 

CONC_OUT_1

A1

and level of the output.

 

REC OFS: Allows them input of a DC offset to let the user

 

 

 

manually adjust the output level.

7.4

 

 

AUTO CAL: Enables / Disables the AOUT CALIBRATED

 

 

feature.

 

 

 

CALIBRATED: Performs the same calibration as AOUT

 

 

 

CALIBRATED, but on this one channel only.

 

CONC_OUT_2

A2

 Same as for CONC_OUT_1 but for analog channel A2.

 

CONC_OUT_3

A3

 Same as for CONC_OUT_1 but for analog channel A3 but

 

only if either the optional O2 or CO2 sensors are installed.

 

 

 

 

TEST OUTPUT

A4

 Same as for CONC_OUT_1 but for analog channel A4

7.4.6

(TEST CHANNEL).

 

 

 

AIN

N/A

Initiates a calibration of the A-to-D Converter circuit located on

7.4.7

CALIBRATED

the Motherboard.

 

 

1 Any changes made to RANGE or REC_OFS require recalibration of this output.

Teledyne Analytical Instruments

129

Page 129
Image 129