Teledyne GFC 7001E Predicting Failures Using the Test Functions, Function Condition Behavior

Models: GFC 7001E GFC 7001EM

1 321
Download 321 pages 4.48 Kb
Page 263
Image 263

Maintenance

Model GFC7001E Carbon Dioxide Analyzer

12.2.PREDICTING FAILURES USING THE TEST FUNCTIONS

The Test Functions can be used to predict failures by looking at how their values change over time. Initially it may be useful to compare the state of these Test Functions to the values recorded on the printed record of the final calibration performed on your instrument at the factory, P/N 04307. Table 12-3 can be used as a basis for taking action as these values change with time. The internal data acquisition system (iDAS) is a convenient way to record and track these changes. Use APICOM to download and review this data from a remote location.

Table 12-3: Predictive uses for Test Functions

FUNCTION

CONDITION

BEHAVIOR

 

 

 

STABILITY

Zero Cal

Increasing

 

 

 

CO MEAS

Zero Cal

Decreasing

 

 

 

 

 

Increasing

 

Zero Cal

 

 

 

 

 

Decreasing

MR RATIO

 

 

 

 

 

 

 

Increasing

 

Span Cal

 

 

 

 

 

 

Decreasing

 

 

 

 

 

Increasing > 1”

PRES

Sample

 

 

 

 

Decreasing > 1”

 

 

 

PHT DRIVE

Any, but with

 

Bench Temp at

Increasing

 

48°C

 

 

 

 

OFFSET

Zero Cal

Increasing

 

Decreasing

 

 

 

 

 

SLOPE

Span Cal

Increasing

 

Decreasing

 

 

 

 

 

INTERPRETATION

Pneumatic Leaks – instrument & sample system

Detector deteriorating

Source Aging

Detector deteriorating

Optics getting dirty or contaminated

Source Aging

Detector deteriorating

Contaminated zero gas (H2O)

Source Aging

Detector deteriorating

GFC Wheel Leaking

Pneumatic Leaks

Contaminated zero gas (CO)

Source Aging

Pneumatic Leaks – instrument & sample system

Calibration system deteriorating

GFC Wheel Leaking

Source Aging

Calibration system deteriorating

Pneumatic Leak between sample inlet and Sample Cell

Change in sampling manifold

Dirty particulate filter

Pneumatic obstruction between sample inlet and Sample Cell

Obstruction in sampling manifold

Mechanical Connection between IR-Detector and Sample Cell deteriorating

IR-Photodetector deteriorating

See MR Ratio - Zero Cal Decreasing above

See MR Ratio - Zero Cal Increasing above

See MR Ratio - Span Cal Decreasing above

See MR Ratio – Span Cal Increasing above

Teledyne Analytical Instruments

263

Page 263
Image 263
Teledyne GFC 7001E Predicting Failures Using the Test Functions, Predictive uses for Test Functions, Interpretation