Getting Started

Model GFC7001E Carbon Dioxide Analyzer

3.5.1.STARTUP

After the electrical and pneumatic connections are made, turn on the instrument. The pump and exhaust fan should start immediately. The display should immediately display a single, horizontal dash in the upper left corner of the display. This will last approximately 30 seconds while the CPU loads the operating system.

Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data. During this process, astring of messages will appear on the analyzer’s front panel display.

System waits 3 seconds then

automatically begins its

initialization routine.

.

System is checking the validity and functionality of the Terminus

memory and firmware.

If at this point,

appears, contact Teledyne

Instruments customer service.

The instrument is loading

configuration and calibration data

from the flash chip.

The instrument is loading the

system firmware.

The startup process may hesitate at this point if:

·The Ethernet option is installed;

· DHCP mode is turned on and;

·The instrument is not connected to a functioning network.

The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and start monitoring CO gas.

Teledyne Analytical Instruments

47

Page 47
Image 47
Teledyne GFC 7001EM operation manual Startup