Specifications

Model GFC7001E Carbon Dioxide Analyzer

2.SPECIFICATIONS AND APPROVALS

2.1.SPECIFICATIONS

Table 2-1: M 300E/300EM Basic Unit Specifications

 

Ranges

GFC 7001E: Min: 0-1 ppm; Max: 0-1000 ppm of Full Scale (User selectable)

 

 

 

GFC 7001EM: Min: 0-5 ppm; Max: 0-5000 ppm of Full Scale (User selectable)

 

 

 

 

 

 

Measurement Units

GFC 7001E: ppb, ppm, µg/m3, mg/m3 (user selectable)

 

 

 

GFC 7001EM: ppm, mg/m3 (user selectable)

 

 

 

 

Zero Noise

GFC 7001E: < 0.02 ppm RMS1;

GFC 7001EM: ≤ 0.1 ppm RMS

 

 

Span Noise

GFC 7001E:<0.5% of rdg RMS over 5ppm1, 3;

GFC 7001EM:>0.5% of rdg RMS

 

 

 

over 20ppm

 

 

 

 

 

 

Lower Detectable Limit1

GFC 7001E: < 0.04 ppm;

GFC 7001EM: 0.2 ppm

 

 

 

 

Zero Drift (24 hours) 2

GFC 7001E: < 0.1 ppm;

GFC 7001EM: <0.5 ppm

 

 

 

 

Zero Drift (7 days) 2

GFC 7001E: < 0.2 ppm;

GFC 7001EM: <1.0ppm

 

 

 

 

Span Drift (24 hour2s)

The greater of < 0.5% of reading or 0.1ppm (GFC 7001E), 0.5ppm(GFC 7001EM)

 

 

Span Drift (7 days) 2

The greater of < 1% of reading or 0.5ppm (GFC 7001E), 1 ppm(GFC 7001EM)

 

 

Linearity

GFC 7001E: Better than 1% Full Scale5;

 

 

 

 

 

GFC 7001EM: 0 - 3000 ppm: 1% full scale; 3000 - 5000 ppm: 2% full scale

 

 

Precision

GFC 7001E: The greater of 0.5% of reading or 0.2ppm;

 

 

 

GFC 7001EM: The greater of 1.0% of reading or 1ppm

 

 

 

 

 

 

 

 

 

 

Lag Time 1

10 sec1

 

 

 

 

 

 

Rise/Fall Time 1

<60 sec to 95%1

 

 

 

 

 

 

Sample Flow Rate

800 cm3/min. ±10%

 

 

 

 

 

 

 

O2 Sensor option adds 120 cm³/min to total flow though when installed

 

 

Temperature Range

5 - 40C operating, 10 - 40C EPA Equivalency (GFC 7001E only)

 

 

 

 

 

 

 

 

 

Humidity Range

0-95% RH, Non-Condensing

 

 

 

 

 

Temp Coefficient

< 0.05 % per C (minimum 50 ppb/C)

 

 

 

 

Voltage Coefficient

< 0.05 % per V

 

 

 

 

 

 

 

 

 

 

 

 

Dimensions (HxWxD)

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

 

 

 

 

 

 

 

 

 

 

 

 

Weight

50 lb (22.7 kg)

 

 

 

 

 

 

AC Power

100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A),

 

 

 

 

 

220 – 240 V 50/60 Hz (2.5A)

 

 

 

 

 

Environmental Conditions

Installation Category (Over voltage Category) II Pollution Degree 2

 

 

 

 

 

 

 

 

 

Analog Outputs

4 user configurable outputs

 

 

 

 

 

 

 

 

 

 

 

 

Analog Output Ranges

All Outputs: 0.1V, 1V, 5V or 10V

 

 

 

 

 

 

Three outputs convertible to 4-20 mA isolated current loop.

 

 

 

All Ranges with 5% under/over-range

 

 

 

 

Analog Output Resolution

1 part in 4096 of selected full-scale voltage

 

 

 

 

 

 

 

 

 

 

 

Status Outputs

8 Status outputs from opto-isolators

 

 

 

 

 

 

 

 

 

 

 

 

Control Inputs

6 Control Inputs, 2 defined, 4 spare

 

 

 

 

 

Serial I/O

One (1) RS-232/optional multidrop; One (1) RS-232/optional RS-485 (2 connecters in

 

 

 

parallel)

 

 

 

 

 

 

 

Baud Rate : 300 - 115200

 

 

 

 

 

 

Alarm outputs (optional)

2 opto-isolated alarm outputs and 2 dry contact alarm outputs

 

 

 

 

 

 

 

 

Certifications

USEPA: Reference Method Number EQOA-0992-087

 

 

 

 

 

CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A

 

 

 

 

 

 

 

 

 

 

 