Getting Started

Model GFC7001E Carbon Dioxide Analyzer

3.5.2.WARM UP

The GFC 7001E/EM requires about 60 minutes warm-up time before reliable CO measurements can be taken. During that time, various portions of the instrument’s front panel will behave as shown in Table 3-8. See Figure 3-1 for the layout.

Table 3-8: Front Panel Display during System Warm-Up

NAME

COLOR

BEHAVIOR

SIGNIFICANCE

Concentration

 

Displays current,

 

N/A

compensated CO

This is normal operation.

Field

 

Concentration

 

 

 

 

 

 

 

 

Mode Field

N/A

Displays blinking

Instrument is in sample mode but is still in the process of

“SAMPLE”

warming up. (iDAS holdoff period is active)

 

 

STATUS LED’s

 

 

 

 

 

 

Unit is operating in sample mode; front panel display is being

Sample

Green

On

updated.

 

 

 

Flashes On/Off when adaptive filter is active

Cal

Yellow

Off

The instrument’s calibration is not enabled.

 

 

 

The analyzer is warming up and hence out of specification

Fault

Red

Blinking

for a fault-free reading. Various warning messages will

 

 

 

appear.

3.5.3.WARNING MESSAGES

Because internal temperatures and other conditions may be outside the specified limits during the analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes after power up. If warning messages persist after the 60 minutes warm-up period is over, investigate their cause using the troubleshooting guidelines in Section Error! Reference source not found..

To view and clear warning messages, press:

Table 3-6 lists brief descriptions of the warning messages that may occur during start up.