Calibration Procedures

Model GFC7001E Carbon Dioxide Analyzer

9.7.CALIBRATION OF OPTIONAL SENSORS

9.7.1.O2 SENSOR CALIBRATION PROCEDURE

9.7.1.1.O2 Calibration Setup

The pneumatic connections for calibrating are as follows:

Figure 9-7: O2 Sensor Calibration Set Up

O2 SENSOR ZERO GAS: Teledyne recommends using pure N2 when calibration the zero point of your O2 sensor option.

O2 SENSOR SPAN GAS: Teledyne recommends using 20.8% O2 in N2 when calibration the span point of your O2 sensor option (See Table 3-7).

Teledyne Analytical Instruments

209

Page 209
Image 209
Teledyne GFC 7001E Calibration of Optional Sensors, 1. O2 Sensor Calibration Procedure, 1.1. O2 Calibration Setup