Optional Hardware and Software

Model GFC7001E Carbon Dioxide Analyzer

5.6.3.ZERO/SPAN/SHUTOFF VALVE (OPTION 50B)

This option requires that both zero air and span gas be supplied from external sources.

Span gas will be supplied from a pressurized bottle of calibrated CO gas.

A critical flow control orifice, internal to the instrument ensures that the proper flow rate is maintained.

An internal vent line ensures that the gas pressure of the span gas is reduced to ambient atmospheric pressure.

A SHUTOFF valve preserves the span gas source when it is not in use.

Zero gas is supplied by either an external scrubber or a zero air generator such as the Teledyne’s M701.

5.6.3.1.Internal Pneumatics (OPT 50B)

Figure 5-5:

Internal Pneumatic Flow OPT 50B – Zero/Span/Shutoff Valves

 

Table 5-2: Zero/Span Valve Operating States for Option 50B

 

 

 

 

 

 

 

MODE

 

VALVE

CONDITION

 

 

SAMPLE

Sample/Cal

Open to SAMPLE inlet

 

 

 

 

 

 

(Normal

 

Zero/Span

Open to IZS inlet

 

 

State)

 

 

 

 

 

 

Shutoff Valve

Closed

 

 

 

 

 

 

 

 

Sample/Cal

Open to ZERO/SPAN valve

 

 

ZERO CAL

Zero/Span

Open to IZS inlet

 

 

 

 

Shutoff Valve

Closed

 

 

 

 

Sample/Cal

Open to ZERO/SPAN valve

 

 

SPAN CAL

 

 

 

 

Zero/Span

Open to SHUTOFF valve

 

 

 

 

Shutoff Valve

Open to PRESSURE SPAN Inlet

 

 

 

 

 

 

 

Teledyne Analytical Instruments

67

Page 67
Image 67
Teledyne GFC 7001EM operation manual ZERO/SPAN/SHUTOFF Valve Option 50B, Internal Pneumatics OPT 50B