Teledyne GFC 7001EM operation manual Repair Procedures, Repairing Sample Flow Control Assembly

Models: GFC 7001E GFC 7001EM

1 321
Download 321 pages 4.48 Kb
Page 299
Image 299

Troubleshooting & Repair

Model GFC7001E Carbon Dioxide Analyzer

13.6.REPAIR PROCEDURES

This contains procedures that might need to be performed on rare occasions when a major component of the analyzer requires repair or replacement.

13.6.1.REPAIRING SAMPLE FLOW CONTROL ASSEMBLY

The critical flow orifice is housed in the flow control assembly (Teledyne P/N 001760400) located on the top of the optical bench. A sintered filter protects the jewel orifice so it is unusual for the orifice to need replacing, but if it does, or the filter needs replacement please use the following procedure (see the Spare Parts list in Appendix B for part numbers and kits):

1.Turn off power to the analyzer.

2.Locate the assembly attached to the sample pump. See Figure 3-4.

3.Disconnect the pneumatic connection from the flow assembly and the assembly from the pump.

4.Remove the fitting and the components as shown in the exploded view below.

5.Replace the o-rings (P/N OR0000001) and the sintered filter (P/N FL0000001).

6.If replacing the critical flow orifice itself (P/N 000941000), make sure that the side with the colored window (usually red) is facing upstream to the flow gas flow.

7.Apply new Teflon® tape to the male connector threads.

8.Re-assemble in reverse order.

9.After reconnecting the power and pneumatic lines, flow check the instrument as described in Section 12.3.4.

Pneumatic Connector, Male 1/8” (P/N FT_70

Spring

(P/N HW_20)

Sintered Filter (P/N FL_01)

Critical Flow Orifice (P/N 00094100)

Make sure it is placed with the jewel down)

O-Ring (P/N OR_01)

Purge Housing (P/N 000850000)

Figure 13-14: Critical Flow Restrictor Assembly Disassembly

Teledyne Analytical Instruments

299

Page 299
Image 299
Teledyne GFC 7001EM operation manual Repair Procedures, Repairing Sample Flow Control Assembly