Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

The iDAS can be disabled, as opposed to suspended, only by disabling or deleting its individual data channels.

7.1.2.IDAS STRUCTURE

The iDAS is designed around the feature of a “record”. A record is a single data point. The type of data recorded in a record is defined by two properties:

PARAMETER type that defines the kind of data to be stored (e.g. the average of gas concentrations measured with three digits of precision). See Section 7.1.5.3.

A TRIGGER event that defines when the record is made (e.g. timer; every time a calibration is performed, etc.). See Section 7.1.5.2.

The specific PARAMETERS and TRIGGER events that describe an individual record are defined in a construct called a DATA CHANNEL (see Section 7.1.3). Each data channel is related one or more parameters with a specific trigger event and various other operational characteristics related to the records being made (e.g. the channels name, number or records to be made, time period between records, whether or not the record is exported via the analyzer’s RS-232 port, etc.).

7.1.2.1.iDAS Channels

The key to the flexibility of the iDAS is its ability to store a large number of combinations of triggering events and data parameters in the form of data channels. Users may create up to 50 data channels and each channel can contain one or more parameters. For each channel, the following are selected:

One triggering event is selected.

Up to 50 data parameters, which can be the shared between channels.

Several other properties that define the structure of the channel and allow the user to make operational decisions regarding the channel.

Table 7-2: iDAS Data Channel Properties

 

PROPERTY

DESCRIPTION

DEFAULT

SETTING RANGE

 

 

 

SETTING

 

 

 

 

 

 

 

 

NAME

The name of the data channel.

“NONE”

Up to 6 letters or digits 1.

 

 

 

TRIGGERING

The event that triggers the data channel to measure

ATIMER

Any available event

 

 

 

EVENT

and store the datum.

(see Appendix A-5).

 

 

 

 

 

 

NUMBER AND

A User-configurable list of data types to be

1

Any available parameter

 

 

 

LIST OF

 

 

recorded in any given channel.

(COMEAS)

(see Appendix A-5).

 

 

 

PARAMETERS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The amount of time between each channel data

000:01:00

000:00:01 to

 

 

 

REPORT PERIOD

366:23:59

 

 

 

 

point.

(1 hour)

(Days:Hours:Minutes)

 

 

 

 

 

 

 

 

NUMBER OF

The number of reports that will be stored in the data

100

1 to 1 million, limited by

 

 

 

file. Once the limit is exceeded, the oldest data is

 

 

RECORDS

available storage space.

 

 

 

over-written.

 

 

 

 

 

 

 

 

 

RS-232 REPORT

Enables the analyzer to automatically report

OFF

OFF or ON

 

 

 

channel values to the RS-232 ports.

 

 

 

 

 

 

 

 

CHANNEL

Enables or disables the channel. Allows a channel

ON

OFF or ON

 

 

 

ENABLED

to be temporarily turned off without deleting it.

 

 

 

 

 

 

 

CAL HOLD OFF

Disables sampling of data parameters while

OFF

OFF or ON

 

 

 

instrument is in calibration mode 2.

 

 

1 More with APICOM, but only the first six are displayed on the front panel.

 

 

 

 

 

2 When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode. DAS HOLD OFF SET in

 

 

the VARS menu (see Section Error! Reference source not found.).

 

 

 

 

 

 

 

 

 

 

 

 

 

Teledyne Analytical Instruments

108

Page 108
Image 108
Teledyne GFC 7001EM operation manual Idas Structure, IDAS Channels, IDAS Data Channel Properties