Troubleshooting & Repair

Model GFC7001E Carbon Dioxide Analyzer

13.6.2.REMOVING/REPLACING THE GFC WHEEL

When removing or replacing the GFC Wheel it is important to perform the disassembly in the following order to avoid damaging the components:

1.Turn off the analyzer.

2.Remove the top cover.

3.Open the instrument’s hinged front panel.

4.Locate the GFC Wheel/motor assembly. See Figure 3-4.

5.Unplug the following electronic components:

The GFC Wheel housing temperature sensor

GFC Wheel heater

GFC Wheel motor power supply

SOURCE ASSEMBLY

THERMISTOR

HEATER

SYNCHRONOUS MOTOR

SAFETY SHIELD

Figure 13-15: Opening the GFC Wheel Housing

6.Remove the three (3) screws holding the opto-pickup printed circuit assembly to the GFC Wheel housing.

Teledyne Analytical Instruments

300

Page 300
Image 300
Teledyne GFC 7001EM operation manual REMOVING/REPLACING the GFC Wheel, Opening the GFC Wheel Housing