Maintenance Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 261
Table 12-1: GFC 7001E/EM Maintenance Schedule
ITEM ACTION FREQ CAL CHECK REQ’D MANUALDATE PERFORMED
Particulate
Filter Replace Weekly or As
Needed No
Verify Test
Functions
Record and
Analyze
Weekly or after
any
Maintenance or
Repair
No
Pump
Diaphragm Replace Annually Yes
Perform Flow
Check Check Flow Annually No
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance or
Repair
No
Pneumatic
lines
Examine and
Clean As Needed Yes if
cleaned
Cleaning Clean As Needed
Only if
cover
removed