Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

7.1.5.9.The Compact Report Feature

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five parameters are reported in one line.

The COMPACT DATA REPORT generally cannot be accessed from the standard iDAS front panel menu, but is available via the instruments communication ports by using APICOM or the analyzer’s standard serial data protocol.

7.1.5.10.The Starting Date Feature

This option allows the user to specify a starting date for any given channel in case the user wants to start data acquisition only after a certain time and date. If the STARTING DATE is in the past (the default condition), the iDAS ignores this setting and begins recording data as defined by the REPORT PERIOD setting.

The STARTING DATE generally cannot be accessed from the standard iDAS front panel menu, but is available via the instruments communication ports by using APICOM or the analyzer’s standard serial data protocol.

7.1.6.DISABLING/ENABLING DATA CHANNELS

Data channels can be temporarily disabled, which can reduce the read/write wear on the Disk-on-Chip. To disable a data channel, follow the instruction shown in Section 7.1.5 then press: