Remote Operation

Model GFC7001E Carbon Dioxide Analyzer

8.6.7.SETTING HESSEN PROTOCOL STATUS FLAGS

Teledyne’s implementation of Hessen protocols includes a set of status bits that the instrument includes in responses to inform the host computer of its condition. Each bit can be assigned to one operational and warning message flag. The default settings for these bit/flags are:

Table 8-8: Default Hessen Status Flag Assignments

 

 

STATUS FLAG NAME

 

DEFAULT BIT

 

 

 

 

ASSIGNMENT

 

 

 

 

 

 

 

 

 

 

 

 

WARNING FLAGS

 

 

 

 

 

SAMPLE FLOW WARNING

 

 

 

 

0001

 

 

 

BENCH TEMP WARNING

 

 

 

 

0002

 

 

 

SOURCE WARNING

 

 

 

 

0004

 

 

 

BOX TEMP WARNING

 

 

 

 

0008

 

 

 

WHEEL TEMP WARNING

 

 

 

 

0010

 

 

 

SAMPLE TEMP WARN

 

 

 

 

0020

 

 

 

SAMPLE PRESS WARN

 

 

 

 

0040

 

 

 

INVALID CONC

 

 

 

 

0080

 

 

 

(The Instrument’s Front Panel Display Will Show The

 

 

 

 

 

Concentration As “Warnings”)

 

 

 

 

 

 

 

 

 

OPERATIONAL FLAGS1

 

 

 

 

 

 

Instrument OFF

 

 

 

 

0100

 

 

 

In MANUAL Calibration Mode

 

 

 

 

0200

 

 

 

In ZERO Calibration Mode4

 

 

 

 

0400

 

 

 

In O2 Calibration Mode (if O2 sensor installed )2,4

 

 

0400

 

 

 

In CO2 Calibration Mode (if CO2 sensor installed )2,4

 

 

0400

 

 

 

In SPAN Calibration Mode

 

 

 

 

0800

 

 

 

UNITS OF MEASURE FLAGS

 

 

 

 

 

 

 

 

UGM

 

 

 

 

0000

 

 

 

MGM

 

 

 

 

2000

 

 

 

PPB

 

 

 

 

4000

 

 

 

PPM

 

 

 

 

6000

 

 

 

SPARE/UNUSED BITS

 

 

 

 

1000, 8000

 

 

 

 

UNASSIGNED FLAGS (0000)

 

 

 

 

 

AZERO WARN2

 

DCPS WARNING

 

 

 

 

 

CANNOT DYN SPAN2

 

REAR BOARD NOT DET

 

 

CANNOT DYN ZERO3

 

SYNC WARNING1

 

 

 

 

 

CONC ALARM 13

 

SYSTEM RESET1

 

 

 

 

 

CONC ALARM 23

 

 

 

 

 

 

 

 

1

These status flags are standard for all instruments and should probably not be

 

 

2

modified.

 

 

 

 

 

 

 

 

Only applicable if the optional internal span gas generator is installed.

 

 

3

Only applicable if the analyzer is equipped with an alarm options.

 

 

3

It is possible to assign more than one flag to the same Hessen status bit. This

 

 

 

allows the grouping of similar flags, such as all temperature warnings, under the

 

 

 

same status bit.

 

 

 

 

 

 

 

 

 

Be careful not to assign conflicting flags to the same bit as each status bit will be

 

 

 

triggered if any of the assigned flags is active.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 