Advanced Features

Model GFC7001E Carbon Dioxide Analyzer

7.2.SETUP MORE VARS: INTERNAL VARIABLES (VARS)

The GFC 7001E/EM has several user-adjustable software variables, which define certain operational parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be manually redefined using the VARS menu.

The following table lists all variables that are available within the 818 password protected level. See Appendix A-2 for a detailed listing of all of the GFC 7001E/EM variables that are accessible through the remote interface.

Table 7-4: Variable Names (VARS)

 

 

 

 

 

ALLOWED

VARS

NO.

VARIABLE

 

DESCRIPTION

DEFAULT

 

VALUES

 

 

 

 

 

VALUES

 

 

 

 

 

 

 

 

 

 

Changes the Internal Data Acquisition System (iDAS)

May be set for

 

 

 

 

 

HOLDOFF timer.

 

 

 

 

 

intervals

 

 

0

 

 

No data is stored in the iDAS channels during situations

 

 

DAS_HOLD_OFF

between

15 min.

 

when the software considers the data to be questionable

 

 

 

 

0.5 – 20 min

 

 

 

 

 

such as during warm-up or just after the instrument returns

 

 

 

 

 

 

 

 

 

 

 

from one of its calibration modes to SAMPLE Mode.

 

 

 

1

CONC_PRECISION

Allows the user to set the number of significant digits to the

AUTO, 1, 2,

AUTO

 

right of the decimal point display of concentration and stability

 

3, 4

 

 

 

 

values.

 

 

 

2

DYN_ZERO 1

Dynamic zero automatically adjusts offset and slope of the

 

OFF

 

CO response when performing a zero point calibration during

ON/OFF

 

 

 

 

an AutoCal (see Section 9.4).

 

 

 

3

DYN_SPAN 1

Dynamic span automatically adjusts the offsets and slopes of

 

OFF

 

the CO response when performing a slope calibration during

ON/OFF

 

 

 

 

an AutoCal (see Section 9.4).

 

 

 

4

CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the +

-60 to +60

0 sec

 

sign if the clock is too slow, choose the - sign if the clock is

 

s/day

 

 

 

 

too fast.

 

 

 

5

STABIL_GAS

2

Selects which gas measurement is displayed when the STABIL

CO; CO2 & O2

CO

 

 

test function is selected.

1

Use of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency.

 

2

This VARS only appears if either the optional O2 or CO2 sensors are installed.

 

 

Teledyne Analytical Instruments

125