Optional Hardware and Software

Model GFC7001E Carbon Dioxide Analyzer

5.6.4.2.Pneumatic Set Up (OPT 50H)

See Figure 3-2 for the location of gas inlets and outlets and span gas no shutoff valves are required.

Figure 5-8: Pneumatic Connections – Option 50H: Zero/Span Calibration Valves

SAMPLE GAS SOURCE:

Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long.

Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-Hg above ambient pressure and ideally should equal ambient atmospheric pressure.

In applications where the sample gas is received from a pressurized manifold, a vent must be placed on the sample gas before it enters the analyzer.

CALIBRATION GAS SOURCES:

SPAN GAS:

Attach a gas line from the source of calibration gas (e.g. a Teledyne’s M700E Dynamic Dilution Calibrator) to the SPAN inlet.

ZERO AIR:

Zero air is supplied internally via a zero air scrubber that draws ambient air through the IZS inlet.

INPUT GAS VENTING:

The zero air supply and sample gas line MUST be vented in order to ensure that the gases input do not exceed the maximum inlet pressure of the analyzer as well as to prevent back diffusion and pressure effects. These vents should be:

At least 0.2m long;

No more than 2m long and;

Vented outside the shelter or immediate area surrounding the instrument.

EXHAUST OUITLET

Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line should be:

PTEF tubing; minimum O.D ¼”;

A maximum of 10 meters long;

Vented outside the analyzer’s enclosure.

Teledyne Analytical Instruments

70

Page 70
Image 70
Teledyne GFC 7001EM operation manual Pneumatic Set Up OPT 50H, Exhaust Ouitlet