Maintenance

Model GFC7001E Carbon Dioxide Analyzer

12.MAINTENANCE SCHEDULE & PROCEDURES

Predictive diagnostic functions, including data acquisition records, failure warnings and test functions built into the analyzer, allow the user to determine when repairs are necessary without performing painstaking preventative maintenance procedures. There are, however, a minimal number of simple procedures that when performed regularly will ensure that the analyzer continues to operate accurately and reliably over its lifetime. Repairs and troubleshooting are covered in Section Error! Reference source not found. of this manual.

12.1.MAINTENANCE SCHEDULE

Table 12-1 shows a typical maintenance schedule for the analyzer. Please note that in certain environments (i.e. dusty, very high ambient pollutant levels) some maintenance procedures may need to be performed more often than shown.

NOTE

A Span and Zero Calibration Check (see CAL CHECK REQ’D Column of Table 12-1) must be performed

following certain of the maintenance procedure listed below. See Sections 9.3 and 9.4 for instructions on performing checks.

CAUTION

GENERAL SAFETY HAZARD

Risk of electrical shock. Disconnect power before performing any of the following

operations that require entry into the interior of the analyzer.

CAUTION

QUALIFIED PERSONNEL

The operations outlined in this section are to be performed by qualified maintenance

personnel only.

Teledyne Analytical Instruments

259

Page 259
Image 259
Teledyne GFC 7001EM operation manual Maintenance Schedule & Procedures, Qualified Personnel