7.4.3.4.Manual Calibration of the Analog Outputs Configured for Voltage Ranges

For highest accuracy, the voltages of the analog outputs can be manually calibrated.

NOTE:

The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL feature

is turned off for the channel being adjusted (see Section 7.4.3.1).

Calibration is performed with a voltmeter connected across the output terminals and by changing the actual output signal level using the front panel keys in 100, 10 or 1 count increments. See Figure 3-7 for pin assignments and diagram of the analog output connector.

V

+DC Gnd

Figure 7-5: Setup for Checking / Calibrating DCV Analog Output Signal Levels

Table 7-8: Voltage Tolerances for the TEST CHANNEL Calibration

FULL

ZERO

 

SPAN

MINIMUM

SPAN VOLTAGE

ADJUSTMENT

SCALE

TOLERANCE

TOLERANCE

 

(1 count)

 

 

 

 

 

 

 

 

 

0.1 VDC

±0.0005V

90 mV

±0.001V

0.02 mV

 

 

 

 

 

1 VDC

±0.001V

900 mV

±0.001V

0.24 mV

 

 

 

 

 

5 VDC

±0.002V

4500 mV

±0.003V

1.22 mV

10 VDC

±0.004V

4500 mV

±0.006V

2.44 mV