Specifications

Model GFC7001E Carbon Dioxide Analyzer

 

 

 

1 As defined by the USEPA

2 At constant temperature and pressure

 

2.2.EPA EQUIVALENCY DESIGNATION

Teledyne’s GFC 7001E Carbon Monoxide Analyzer is designated as Reference Method Number EQOA-0992-

087 as defined in 40 CFR Part 53, when operated under the following conditions:

Range: Any range from 10 ppm to 50 ppm.

Ambient temperature range of 10 to 40C.

Line voltage range of 90 – 127 and 200 – 230 VAC, 50/60 Hz.

Sample filter: Equipped with PTFE filter element in the internal filter assembly.

Sample flow of 800  80 cm3/min at sea level.

Internal sample pump.

Software settings:

Dilution factor

1.0

AutoCal

ON or OFF

Dynamic Zero

ON or OFF

 

 

Dynamic Span

OFF

 

 

Dual range

ON or OFF

 

 

Auto range

ON or OFF

Temp/Pres compensation

ON

Under the designation, the analyzer may be operated with or without the following options:

Rack mount with slides.

Rack mount without slides, ears only.

Zero/span valve options.

Option 50A – Sample/Cal valves, or;

Option 50B – Sample/Cal valves with span shutoff & flow control.

Internal zero/span (IZS) option with either:

Option 51A – Sample/Cal valves, or;

Option 51C – Sample/Cal valves with span shutoff & flow control.

Status outputs.

Control inputs.

RS-232 output.

Ethernet output.

4-20mA, isolated output.