Optional Hardware and Software

Model GFC7001E Carbon Dioxide Analyzer

5.6.5.ZERO/SPAN/SHUTOFF WITH INTERNAL ZERO AIR SCRUBBER (OPTION 50E)

5.6.5.1.Internal Pneumatics (OPT 50E)

Figure 5-9: Internal Pneumatic Flow OPT 50E – Zero/Span/Shutoff Valves with Internal Zero Air

Scrubber

Table 5-4: Zero/Span Valve Operating States for Option 50E

Mode

Valve

Condition

 

 

 

SAMPLE

Sample/Cal

Open to SAMPLE inlet

(Normal

Zero/Span

Open to internal ZERO AIR scrubber

State)

 

 

Shutoff Valve

Closed

 

 

Sample/Cal

Open to zero/span valve

ZERO CAL

Zero/Span

Open to internal ZERO AIR scrubber

 

Shutoff Valve

Closed

 

 

 

 

Sample/Cal

Open to ZERO/SPAN valve

SPAN CAL

 

 

Zero/Span

Open to SHUTOFF valve

 

Shutoff Valve

Open to PRESSURE SPAN inlet

Teledyne Analytical Instruments

71

Page 71
Image 71
Teledyne GFC 7001EM ZERO/SPAN/SHUTOFF with Internal Zero AIR Scrubber Option 50E, Internal Pneumatics OPT 50E