Basic Operation

Model GFC7001E Carbon Dioxide Analyzer

6.2.SAMPLE MODE

This is the analyzer’s standard operating mode. In this mode the instrument is analyzing the gas in the sample chamber, calculating CO concentration and reporting this information to the user via the front panel display, the analog outputs and, if set up properly, the RS-232/RS-485/Ethernet ports.

NOTE

A value of “XXXX” displayed in the CO Concentration field means that the M/R ratio is invalid because

CO REF is either too high (> 4950 mVDC) or too low (< 1250 VDC).

A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range reading or the

analyzer’s inability to calculate it.

A variety of TEST functions are available for viewing at the front panel whenever the analyzer is at the MAIN MENU. These functions provide information about the various functional parameters related to the analyzers operation and its measurement of gas concentrations. This information is particularly a performance problem during troubleshooting (see Section 13.1.2).

To view these TEST functions, press,

Figure 6-2: Viewing GFC 7001E/EM Test Functions

NOTE

All pressure measurements are represented in terms of absolute pressure. Absolute, atmospheric

pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and passing storms can cause changes in the absolute atmospheric pressure.

Teledyne Analytical Instruments

86

Page 86
Image 86
Teledyne GFC 7001EM operation manual Sample Mode