Maintenance

Model GFC7001E Carbon Dioxide Analyzer

12.3.MAINTENANCE PROCEDURES

The following procedures are to be performed periodically as part of the standard maintenance of the GFC 7001E.

12.3.1.REPLACING THE SAMPLE PARTICULATE FILTER

The particulate filter should be inspected often for signs of plugging or contamination. We recommend that the filter and the wetted surfaces of the filter housing are handled as little as possible when you change the filter. Do not touch any part of the housing, filter element, PTFE retaining ring, glass cover and the o-ring.

To change the filter:

1.Turn OFF the analyzer to prevent drawing debris into the instrument.

2.Open the GFC 7001E Analyzer’s hinged front panel and unscrew the knurled retaining ring on the filter assembly.

Figure 12-1: Sample Particulate Filter Assembly

3.Carefully remove the retaining ring, PTFE o-ring, glass filter cover and filter element.

4.Replace the filter, being careful that the element is fully seated and centered in the bottom of the holder.

5.Re-install the PTFE o-ring (with the notches up), the glass cover, then screw on the retaining ring and hand tighten. Inspect the seal between the edge of filter and the o-ring to assure a proper seal.

6.Re-start the Analyzer.

12.3.2.REBUILDING THE SAMPLE PUMP

The diaphragm in the sample pump periodically wears out and must be replaced. A sample rebuild kit is available – see label on pump for the part number of the pump rebuild kit. Instructions and diagrams are included with the kit.

Always perform a Flow and Leak Check after rebuilding the Sample Pump.

Teledyne Analytical Instruments

264

Page 264
Image 264
Teledyne GFC 7001EM Maintenance Procedures, Replacing the Sample Particulate Filter, Rebuilding the Sample Pump