Getting Started

Model GFC7001E Carbon Dioxide Analyzer

3.3.4.CONNECTING THE CONTROL INPUTS

If you wish to use the analyzer to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.

There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method. However, if full isolation is required, an external 5 VDC power supply should be used.

 

CONTROL IN

 

 

CONTROL IN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A B C D E F

ZERO

SPAN

RANGE

 

LOW

HIGH

 

 

CAL

U +

A B

C D E F

 

 

ZERO

SPAN

RANGE

 

 

 

 

 

 

 

LOW

HIGH

 

 

 

 

CAL

U+

-

5 VDC Power

Supply

+

 

 

 

Local Power Connections

 

External Power Connections

 

 

 

 

 

Figure 3-9:

Control Input Connector

 

 

 

 

 

Table 3-6: Control Input Signals

 

 

 

 

 

 

 

 

 

 

 

 

INPUT #

STATUS DEFINITION

 

 

ON CONDITION

 

 

 

A

REMOTE ZERO CAL

 

The analyzer is placed in Zero Calibration mode. The mode field of the

 

 

 

 

display will read ZERO CAL R.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The analyzer is placed in span calibration mode as part of performing a low

 

B

REMOTE SPAN CAL

 

span (midpoint) calibration. The mode field of the display will read LO CAL

 

 

 

 

 

R.

 

 

 

 

 

 

REMOTE CAL HIGH

 

The analyzer is forced into high range for zero or span calibrations. This

 

 

 

C

 

only applies when the range mode is either DUAL or AUTO. The mode field

 

RANGE

 

 

 

 

 

of the display will read HI CAL R.

 

 

 

 

 

 

 

 

 

 

D, E

SPARE

 

 

 

 

 

& F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Digital Ground

 

The ground level from the analyzer’s internal DC power supplies (same as

 

 

 

 

 

 

 

 

chassis ground).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

U

External Power input

 

Input pin for +5 VDC required to activate pins A – F.

 

 

 

 

 

5 VDC output

 

Internally generated 5V DC power. To activate inputs A – F, place a jumper

 

+

 

between this pin and the “U” pin. The maximum amperage through this port

 

 

 

 

 

is 300 mA (combined with the analog output supply, if used).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Teledyne Analytical Instruments

41

Page 41
Image 41