TROUBLESHOOTING | SECTION 2 |
|
|
Once the motor brush caps have been removed, pull the brushes out of the motor. The end of the brushes should be smooth and shiny and the spring should not be damaged or discolored. If one or both of the brushes are damaged, only the damaged or worn brushes need be replaced. It is very important that any time a brush is replaced, it must be “burned in”. This is accomplished by running the motor for one hour in each direction with a half hour break
ELECTRO-MECHANICAL PARKING BRAKE TESTING (FIGURE 4)
NOTE: This procedure should only be performed on wheelchairs with conventional motor/ gearbox assembly.
1.On the
2.Set the digital multimeter to read ohms (Ω).
3.Measure the resistance between the two (2) brake contacts. A normal reading is 100 ohms (Ω). A reading of 0 ohms (Ω) or a very high reading; i.e., MEG ohms or O.L. (out of limit) indicates a shorted brake or an open connection respectively. If either condition exists, send the motor to Invacare Technical Service for inspection/repair.
TROUBLESHOOTING
CAUTION
A shorted
NOTE: A bad motor can damage the controller module but a bad controller will NOT damage a motor.
4 Pin Motor Connector
Ohmmeter
FIGURE 4 - ELECTRO-
MECHANICAL PARKING BRAKE
TESTING
Part No. 1104849 | 19 | 3G Storm Series® Wheelchairs |