RECLINER

SECTION 14

 

 

REPLACING/ADJUSTING GAS CYLINDERS (FIGURE 4)

WARNING

Replace ONE (1) gas cylinder at a time to avoid injury.

Both gas cylinders MUST be operational and adjusted properly BEFORE using the recliner. DO NOT operate the recliner if only one (1) of the gas cylinders is operational or adjusted properly.

REPLACING GAS CYLINDER

1.Remove the mounting screw, washer, nylon washers and locknut that secure the TOP ofRthe gas cylinder to

mounting bracketE on back cane.

C

2.Loosen theLjam nut on existing gas cylinder rodI .

3.Unscrew theN existing gas cylinder

E

from the operator of the recliner

R

cable assembly and the pivot block and remove the existing gas cylinder from the wheelchair.

4.Screw cylinder rod of the NEW gas cylinder into the pivot block until the threads of the cylinder rod are flush with inside of pivot block (FIGURE 4).

5.With the operator of the recliner cable assembly on the inside of the recliner seat frame, line up the mounting hole in the operator of the recliner cable assembly with the new gas cylinder rod.

Nylon Washer

Mounting Bracket

Handle

Mounting Screw

 

 

 

 

Washer

 

 

Lock Nut

 

Gas Cylinder

 

Rod

No Slack

 

in Cable

 

Jam Nut

 

 

 

Gas Cylinder

 

Gas

Rod Threads

 

Cylinder

Pivot Block

 

 

 

Cable Snapped

NO Movement in

into Slot

Operator of the Recliner

 

 

Cable Assembly

 

Cable Fitting

Handle

 

Seated Properly

FIGURE 4 - REPLACING/

ADJUSTING GAS CYLINDERS

RECLINER

CAUTION

DO NOT force the gas cylinder rod into the operator of the recliner cable assembly.

DO NOT cross thread the operator of the recliner cable assembly with the gas cylinder.

If slack in the recliner cable or movement in the operator of the cable assembly can not be eliminated, DO NOT use the recliner cable assembly.

6.Screw the NEW gas cylinder into the operator of the cable assembly until the jam nut sits on the pivot block, there is no slack in recliner cable and there is no movement in the operator of the recliner cable assembly.

7.Visually inspect the handle to make sure that the cable is snapped completely into slot in handle and cable fitting is seat properly in the handle.

8.Press the operator of the recliner cable assembly to extend the NEW gas cylinder.

Part No. 1104849