Sigma SQC-122c manual Chapter Operation

Page 30

Chapter 2

Operation

about 10% overshoot, lower the P Term. If the time to reach PWRDR is very slow, increase the P Term. A lower I Term will increase response time, a higher value will eliminate ringing and setpoint deviations. It is unlikely you will need any D Term.

Continue to Start the process and adjust PID until steady-state response is smooth and the step response is reasonably controlled. You don’t need to totally eliminate ringing during the step if the steady-state response is smooth, preconditioning will minimize step changes.

Set Preconditioning: The power level you recorded as PWR0R is the power where deposition just begins. That’s a good value for Ramp 1 power. PWRDR, or slightly less, is a good value for Ramp 2 Power. This will eliminate a large step change when entering the deposition phase.

Once PID terms are established for a material, they will typically be similar for other materials. Only the P Term and preconditioning power levels may need adjustment.

2-16

Image 30
Contents SQC-122c Thin Film Deposition Controller User’s Guide Safety Information Safety SymbolsWarranty Information Limitation of WarrantyReturn Policy Chapter Menus Table of Contents Chapter Quick StartChapter Operation Chapter OptionsAppendix Chapter Quick Start Introduction Thin Film Process OverviewChapter Quick Start Front Panel Power Input Fuse Chapter Quick Start Rear PanelChapter Quick Start System Connections Chapter Quick Start Installation Quick Setup Chapter Quick Start Process SetupPower SelectEdit Mode Layer Chapter Quick StartExit to Main Chapter Quick Start Depositing a Film This page left blank for your notes Chapter Operation Introduction DefinitionsDefining a Film POWER% POWER% Deposition Controls Menu Chapter Operation Defining a Process Layer Edit MenuChapter Operation Chapter Operation Sensor Setup Tooling U n d e r 1 0 0 % Chapter Operation Chapter Operation Source Setup Chapter Operation Running a Process Chapter Operation Next Menu Zero THICKNESSkA RATEA/s 824 55.3 Chapter Operation Loop Tuning Chapter Operation Chapter Operation Troubleshooting No Readings, or Erratic Readings from SensorsIncorrect Rate or Thickness Measurement Poor Rate StabilityThis page left blank for your notes Chapter Menus Introduction Default, power-up Main Screen display is shown below Chapter Menus Main MenuMenu Quick SetupNext FilmChapter Menus Next Menu Screens Prev Layer Chapter Menus Quick Setup MenuExit to Main Edit Next LayerChapter Menus Chapter Menus Process Menus Main Screen Prev Menu Select Edit Delete Control KnobMain Screen Prev Menu Edit Cut Paste Insert Layer Control KnobPOWER% Exit to Main Prev Menu Edit Control KnobChapter Menus Chapter Menus Film Menus Main Screen Prev Menu Edit Delete Control KnobExit to Main Prev Menu Edit Exit to Main Prev Menu Edit THICKNESSkA RATEA/s Deposition Controls Menu Chapter Menus Chapter Menus Chapter Menus Chapter Menus System Parameters Menu Chapter Menus Chapter Menus Chapter Menus I/O Setup Mapping MenuRelays Explanation closes when Inputs Explanation high input causesThis page left blank for your notes Full Rack Extender Installation Chapter Options IntroductionOption Card Appendix Material Parameters Measurement Appendix SpecificationsGeneral Specifications SourceDisplay Connector Wiring Relay Pins Input PinsAppendix O Connections Appendix Appendix Handheld Remote Controller Declares that the product Complies with the Essential Health and Safety RequirementsAppendix EC Declaration of Conformity Uses the following standards