Chapter 2 | Operation |
When the frequency reading is stable, reconnect the source supply. Start the deposition process in Manual mode with 0% output. The % Life readings should remain stable.
Slowly raise the % Output until a rate reading is displayed above the graph. As material is deposited on the crystal, the % Life reading should remain stable, or drop slowly and consistently. If not, check your source supply for erratic output. Also assure that the sensor is not too close to the source (particularly in sputtering).
2.8.2Incorrect Rate or Thickness Measurement:
First, complete the procedures in Section 2.8.1 to assure reliable sensor operation.
Set the Xtal Tooling as described in the System Menu section of Chapter 3. Incorrect Xtal Tooling values will cause consistently low or high rate/thickness values for every material.
Once the Xtal Tooling is set, set Film Tooling in the Film Menu to 100 unless you are certain that another value is needed for a specific film.
Verify that the Density and
2.8.3Poor Rate Stability:
First, be sure that a stable rate can be achieved in Manual mode, as explained in Section 2.8.1. Once a stable rate is achieved in Manual mode, follow the Loop Tuning procedures of Section 2.7.