2.7Loop Tuning
This section will help you adjust your control loop PID parameters to achieve a stable deposition process. Keep in mind that there is no “best” way to determine PID parameters, and no one set of settings that are “best.”
Setup System Parameters: Be sure that the output Scale and crystal Min/Max Frequency parameters are accurate for your system. All Tooling parameters are best set to 100% for now. A Period of .25 seconds is also a good starting point. Simulate should be OFF.
Create a
Test the Setup: Press Auto/Manual to start the layer in Manual mode. Slowly turn the control knob to a power of 10%, and verify that your power supply output is about 10% of full scale. Continue to turn the control knob until a Rate(A/s) above 0 is shown.
Again, verify that the power supply output agrees with the
Determine Open Loop Gain: Slowly adjust the control knob until the Rate(A/s) reading approximately matches your desired rate (Init Rate in the Quick Setup menu). Record the desired rate Power(%) reading as PWRDR. Slowly lower the power until the Rate(A/s) reading is just at (or near) zero. Record the zero rate Power(%) reading as PWR0R.
Determine Open Loop Response Time: Calculate 1/3 of your desired rate (RATE1/3), and 2/3 of the desired rate (RATE2/3) for this layer. Slowly increase the power until Rate(A/s) matches RATE1/3. Get ready to record the loop’s response to an input change. Quickly adjust Power(%) to PWRDR. Measure the time for the Rate (A/s) reading to reach RATE2/3. You may want to do this several times to get an average response time reading. Displaying the Rate graph will also help. Twice the measured time is the step response time, TIMESR. TIMESR is typically .7 to 1.5 seconds for E- Beam evaporation, 5 to 20 seconds for thermal evaporation.
Return the output power to 0%, then press Manual/Auto to return to Auto mode. Follow these steps to set the loop PID parameters:
Set PID Values: In the Quick Setup menu set P=25, I= TIMESR, D=0. Exit the Quick Setup menu. Press Start and observe the Power graph. The power should rise from 0%, and stabilize near PWRDR with little ringing or overshoot. If there is more than